Home · Prototypes · Ethanol Vapor Generator
Automated ethanol vapor generator with pneumatic cylinder, flask, pumps, valves, and hoses on a lab bench.
AUD animal studies · Thermodynamic vapor control

Automated Ethanol Vapor Generator

A controlled ethanol vapor source for Alcohol Use Disorder animal studies, built to provide repeatable vapor concentration from a single governing parameter: temperature under equilibrium conditions.

ApplicationAlcohol vapor inhalation studies for AUD animal models
Control principleEquilibrium ethanol vapor pressure governed by flask temperature
Ethanol flask95% ethanol held at 25 °C
Pneumatic cylinder1.2 L cylinder, heated to 42 °C to prevent condensation
Cycle outputApproximately 3.6 L saturated vapor/air mix in 1 minute
Measured productionApproximately 100 mg/L, about 5.3% v/v vapor-to-air mixture
ValidationStable vapor after two to three priming cycles

Why we built it

Common ethanol vapor systems can require several interacting controls or operate under non-equilibrium conditions that drift over time. This design uses vacuum-assisted evaporation into a pneumatic cylinder and temperature-controlled equilibrium to make the vapor dose more predictable.

How it works

The system evacuates the cylinder, admits saturated ethanol vapor from a temperature-controlled flask, adds air on the cylinder side, then pushes the vapor/air mixture to a downstream mixer or animal cages. Heating the cylinder reduces condensation during expansion and transfer.

Built for research groups that need a reproducible concentrated ethanol vapor/air mixture that can be diluted before animal exposure.

Discuss a vapor system → Read design PDF All prototypes