A controlled ethanol vapor source for Alcohol Use Disorder animal studies, built to provide repeatable vapor concentration from a single governing parameter: temperature under equilibrium conditions.
| Application | Alcohol vapor inhalation studies for AUD animal models |
|---|---|
| Control principle | Equilibrium ethanol vapor pressure governed by flask temperature |
| Ethanol flask | 95% ethanol held at 25 °C |
| Pneumatic cylinder | 1.2 L cylinder, heated to 42 °C to prevent condensation |
| Cycle output | Approximately 3.6 L saturated vapor/air mix in 1 minute |
| Measured production | Approximately 100 mg/L, about 5.3% v/v vapor-to-air mixture |
| Validation | Stable vapor after two to three priming cycles |
Common ethanol vapor systems can require several interacting controls or operate under non-equilibrium conditions that drift over time. This design uses vacuum-assisted evaporation into a pneumatic cylinder and temperature-controlled equilibrium to make the vapor dose more predictable.
The system evacuates the cylinder, admits saturated ethanol vapor from a temperature-controlled flask, adds air on the cylinder side, then pushes the vapor/air mixture to a downstream mixer or animal cages. Heating the cylinder reduces condensation during expansion and transfer.
Built for research groups that need a reproducible concentrated ethanol vapor/air mixture that can be diluted before animal exposure.